Global Global Algérie Français Algeria English Angola Português Angola English Argentina Español Argentina English Australia English Austria Deutsch Austria English Azerbaijan English Azerbaijan Русский Bahrain English Bangladesh English Belgium English Belgium Français Belgium Nederlands Brazil Português Brazil English Bulgaria български Bulgaria English Cameroon English Cameroon Français Canada English Canada Français Chile Español Chile English China 中文 China 日本語 China English Colombia Español Colombia English Croatia English Czech Republic Čeština Czech Republic English Denmark English Djibouti English Ecuador Español Ecuador English Egypt English Estonia English Ethiopia English Finland Suomi Finland English France Français France English Georgia English Georgia Русский Germany Deutsch Germany English Ghana English Greece Ελληνικά Greece English Guatemala Español Guatemala English Hong Kong, China English Hong Kong, China 中文 Hungary Magyar Hungary English India English Indonesia English Iraq English Ireland English Italy Italiano Italy English Ivory Coast Français Ivory Coast English Japan 日本語 Japan English Jordan English Kazakhstan Русский Kazakhstan English Kenya English Kuwait English Kyrgyzstan Русский Kyrgyzstan English Latvia English Lithuania English Malaysia English Mauritius English Mauritius français (Maurice) Mexico Español Mexico English Moldova Română Moldova English Mongolia English Morocco English Morocco Français Mozambique Português Mozambique English Netherlands Nederlands Netherlands English New Zealand English Nigeria English Norway English Oman English Pakistan English Paraguay Español Paraguay English Peru Español Peru English Philippines English Poland Polski Poland English Portugal Português Portugal English Qatar English Republic of Korea 한국어 Republic of Korea English Romania Română Romania English Saudi Arabia English Serbia Српски Serbia English Singapore English Slovakia English Slovenia English South Africa English Spain English Spain Español Sri Lanka English Sweden English Switzerland Deutsch Switzerland Français Switzerland Italiano Switzerland English Tanzania English Thailand ไทย Thailand English Togo English Togo Français Tunisia English Tunisia Français Türkiye Türkçe Türkiye English Turkmenistan Русский Turkmenistan English Ukraine Українська Ukraine English United Arab Emirates English United Kingdom English Uruguay Español Uruguay English USA English Uzbekistan English Uzbekistan Русский Vietnam Tiếng Việt Vietnam English

Hemolysis testing is a core biocompatibility evaluation required for medical devices that contact blood directly or indirectly, quantifying the degree of red blood cell damage caused by a device or its leachables. Conducted in accordance with ISO 10993-4, ASTM F756, and GB/T 16886.4, the test supports regulatory submissions to FDA, EU MDR, and NMPA markets. This article explains the principle, sample requirements, test methods, acceptance criteria, and typical applications of hemolysis testing for medical devices.

Technical Overview

Hemolysis testing is an in vitro biocompatibility test that measures the extent to which a medical device, biomaterial, or its extract damages red blood cells (erythrocytes) and releases hemoglobin into the surrounding medium. If a device surface, residual process chemical, or leachable substance disrupts erythrocyte membranes, free hemoglobin is liberated into plasma, which can clinically manifest as anemia, hemoglobinemia, and kidney injury. Hemolysis assessment is therefore a fundamental safety endpoint for blood-contacting devices.

The test is governed primarily by ISO 10993-4, Biological evaluation of medical devices — Part 4: Selection of tests for interactions with blood, which classifies hemolysis within the hematology category of blood/device interactions, and by ASTM F756, Standard Practice for Assessment of Hemolytic Properties of Materials. In China, the harmonized standard GB/T 16886.4 and the product-level method standard GB/T 14233.1 are commonly referenced by NMPA reviewers for disposable infusion, transfusion, and injection equipment.

ISO 10993-4 assigns hemocompatibility test requirements according to the blood contact category of the device, including:

  • External communicating devices with an indirect blood path, such as IV catheters, syringes, and extension sets
  • External communicating devices with circulating blood, such as hemodialysis circuits, ECMO systems, and blood oxygenators
  • Blood-contacting implant devices, such as stents, vascular grafts, heart valves, and ventricular assist devices

Based on the hemolytic index defined in ASTM F756, materials are typically classified as nonhemolytic (0–2%), slightly hemolytic (2–5%), or hemolytic (greater than 5%). For most regulatory submissions, a hemolysis rate of 5% or less is widely regarded as the acceptance threshold.

Testing Samples

Blood-Contacting Medical Devices

Finished, sterilized devices intended for direct or indirect blood contact are the primary test articles, including intravenous and central venous catheters, hemodialyzers, blood tubing sets, blood collection bags, oxygenators, autotransfusion systems, infusion and transfusion sets, and perfusion cannulae.

Cardiovascular Implants

Stents, stent grafts, vascular grafts, heart valves, septal occluders, pacemaker and defibrillator leads, and ventricular assist device blood-wetted components are evaluated because they reside in contact with blood for extended durations.

Raw Materials and Components

During design and development, polymers such as PVC, polyurethane, and silicone; metals such as stainless steel, nitinol, cobalt-chromium, and titanium; surface coatings; adhesives; and 3D-printed materials are screened for hemolytic potential before device integration.

Sample Preparation Requirements

Samples should represent the final product in its sterilized, packaged condition whenever possible. Extraction media, surface-area-to-volume ratios, and extraction conditions follow ISO 10993-12, for example 3 cm²/mL or 0.2 g/mL ratios, with common extraction conditions such as 37 °C for 72 hours, 50 °C for 72 hours, 70 °C for 24 hours, or 121 °C for 1 hour depending on the device and justification.

Testing Items

The principal test item is the hemolysis rate (hemolytic index), determined by spectrophotometric quantification of free hemoglobin released from diluted anticoagulated blood after contact with the test article or its extract. Typical deliverables include:

  • Hemolysis rate by the direct contact method
  • Hemolysis rate by the extraction (indirect) method
  • Free hemoglobin concentration in the supernatant measured at 540–545 nm
  • Complementary hemocompatibility endpoints under ISO 10993-4, including thrombosis, coagulation (PT, APTT, thrombin time, fibrinogen), platelet count and activation, leukocyte and erythrocyte counts, and complement activation markers such as C3a and SC5b-9

Many biological evaluation programs combine hemolysis testing with these related endpoints to build a complete ISO 10993-4 hemocompatibility data set in a single study at an accredited third-party laboratory.

Testing Methods

Direct Contact Method

Diluted fresh anticoagulated blood, typically rabbit or human blood anticoagulated with sodium citrate, is brought into direct contact with the test article and incubated at 37 °C for a defined period, commonly 2 to 4 hours depending on the protocol applied; GB/T 14233.1 specifies incubation at 37 °C ± 1 °C for 2 hours. After incubation, the blood is centrifuged, and the absorbance of the supernatant is measured at 545 nm to quantify released hemoglobin. This method is preferred when the blood-contacting surface itself is the safety-relevant factor.

Extraction (Indirect) Method

A device extract is first prepared in 0.9% sodium chloride solution or phosphate-buffered saline using the extraction ratios and conditions specified in ISO 10993-12, for example 37 °C for 72 hours. Diluted blood is then incubated with the extract at 37 °C, centrifuged, and the free hemoglobin in the supernatant is quantified spectrophotometrically. This method is suitable when hemolytic risk is driven by leachable substances rather than surface effects.

Controls

Physiological saline or PBS serves as the negative control, producing minimal hemolysis, while sterile distilled or deionized water serves as the positive control, producing complete hemolysis. The difference between positive and negative control absorbance values must meet the validity criteria defined by the test standard before results are accepted.

Calculation and Acceptance Criteria

The hemolysis rate is calculated as: Hemolysis (%) = (Absorbance of sample − Absorbance of negative control) / (Absorbance of positive control − Absorbance of negative control) × 100. A result of 5% or less is generally considered nonhemolytic and acceptable for regulatory purposes, while results above 5% indicate hemolytic potential requiring material or design remediation.

Dynamic Flow Testing

For active blood-handling devices such as blood pumps, oxygenators, and ventricular assist systems, ISO 10993-4 and related ASTM standards for blood pumps describe dynamic in vitro recirculation loop studies in which hemolysis is evaluated under physiologically relevant flow and shear conditions over extended circulation times.

Applications

Hemolysis testing data are used across the entire device lifecycle:

  • Regulatory submissions, including FDA 510(k), De Novo, and PMA applications, EU MDR technical documentation and CE marking, NMPA registration dossiers, and submissions in Japan and Canada
  • Material screening and selection during research and development of blood-contacting polymers, coatings, and metals
  • Design and process change control, such as material supplier changes, sterilization method changes, or manufacturing site transfers
  • Batch release and quality control for infusion, transfusion, and injection equipment in accordance with GB/T 14233.1
  • Biological evaluation plans and toxicological risk assessments under ISO 10993-1 and ISO 14971

For externally communicating and implantable blood-contacting devices, hemolysis testing is almost always expected by regulators and should be planned early in the development program.

FAQ

What hemolysis rate is considered acceptable?

A hemolysis rate of 5% or less is the widely accepted threshold for most regulatory submissions. Under the ASTM F756 classification scheme, materials scoring 0–2% are nonhemolytic, 2–5% are slightly hemolytic, and above 5% are hemolytic.

Should I choose the direct contact method or the extraction method?

The choice depends on how the device contacts blood. Direct contact is appropriate when the device surface interacts with blood, while the extraction method targets hemolytic leachables. Many submissions include both methods to cover surface and extractable effects comprehensively.

What blood is used for the test?

Fresh rabbit blood is most commonly used and is specified in GB/T 14233.1 and ASTM F756 protocols; fresh human donor blood is also acceptable. The blood is anticoagulated, typically with sodium citrate, and used within a defined short period after collection.

Do devices without blood contact require hemolysis testing?

Generally no. The requirement is determined by the blood contact categorization in ISO 10993-1 and ISO 10993-4. Devices with no direct or indirect blood contact can usually be justified as not requiring hemolysis evaluation in the biological evaluation plan.

Which standards should the test report reference?

Reports typically reference ISO 10993-4 for test selection, ISO 10993-12 for extract preparation, ASTM F756 for the hemolytic assessment practice, and GB/T 16886.4 or GB/T 14233.1 for products registered in China, according to the target market.

How many samples are required?

Protocols usually require triplicate test article replicates together with positive and negative controls, using sterile finished devices in their final packaging. Exact sample quantities depend on the device geometry and the extraction ratio applied by the accredited third-party laboratory.

Can hemolysis testing be combined with other biocompatibility tests?

Yes. Hemolysis is frequently bundled with cytotoxicity (ISO 10993-5), sensitization and irritation (ISO 10993-10 and ISO 10993-23), and other hemocompatibility endpoints to form a complete biocompatibility package aligned with the ISO 10993-1 evaluation plan.